Esprit® Ventilator


    Specifications

Esprit Ventilator

Types of Ventilation Modes of Ventilation
Volume Control Assist/Control (A/C)
Pressure Control SIMV
Apnea CPAP
Noninvasive Positive Pressure Spont, Spont/Timed
Percent O2 21 to 100%
Patient Type Adult, Pediatric
Apnea Backup Rate 1 to 80 BPM


Volume Ventilation Settings and Ranges


Tidal Volume 50 ml to 2500 ml
Respiratory Rate 1 to 80 BPM
Plateau 0 to 2.0 sec.
Peak Inspiratory Flow 3 to 140 LPM
(Tubing compliance compensated)
Spontaneous Flow Up to 200 LPM (Uncompensated)
PEEP/CPAP 0 to 35 cm H2O
PSV Pressure 0 to 100 cm H2O
Risetime 0.1 to 0.9 sec.
Percent O2 21 to 100%
Patient Type Adult, Pediatric
Apnea Backup Rate 1 to 80 BPM
Flow Wave Forms Descending, Square
 
Inspiratory Trigger
Pressure Sensitivity -20 to -0.1 cm H2O
Flow Sensitivity 0.5 to 20 LPM
Base Flow 3 LPM above flow sensitivity
(During flow triggering)
Expiratory Trigger
Exhalation Sensitivity 10 to 45% of peak flow


Pressure Control Ventilation Settings and Ranges

Inspiratory Pressure 5 to 100 cm H2O
Inspiratory Time 0.1 to 9.9 sec.
Risetime 0.1 to 0.9 sec.
Respiratory Rate 1 to 80 BPM
PEEP/CPAP 0 to 35 cm H2O
PSV Pressure 0 to 100 cm H2O
Percent O2 21 to 100%
Patient Type Adult, Paediatric
Apnea Backup Rate 1 to 80 BPM
 
Inspiratory Trigger
Pressure Sensitivity -20 to -0.1 cm H2O
Flow Sensitivity 0.5 to 20 LPM
Base Flow 3 LPM above flow sensitivity
(During flow triggering)
Expiratory Trigger
Exhalation Sensitivity 10 to 45% of peak flow


Noninvasive Positive Pressure Ventilation Settings

EPAP/CPAP Pressure 2 to 25 cm H2O
IPAP Pressure 2 to 35 cm H2O
Respiratory Rate 1 to 80 BPM
Inspiratory Time 0.1 to 9.9 sec.
Rise Time 0.1 to 0.9 sec.
Percent O2 21 to 100%
Patient Type Adult, Pediatric
Apnea Backup Rate 1 to 80 BPM
 
Inspiratory Trigger
Flow Sensitivity 0.5 to 20 LPM
Expiratory Trigger
Exhalation Sensitivity 10 to 45% of peak flow


  Front Panel Controls
  • Alarm Reset
  • 100% O2
  • Expiratory Hold
  • Accept
  • Screen Lock
  • Alarm Silence
  • Manual Inspiration
  • Display Brightness
  • Value Knob
  • Alarm Volume Control (Min to max)

Monitored Data and Ranges


Breath Phase and Type Mandatory, Assist,
Supported, Spontaneous
Respiratory Rate (Spontaneous) 0 to 150 BPM
Respiratory Rate (Total) 0 to 150 BPM
Exhaled Tidal Volume 0 to 9999 ml
Exhaled Minute Volume Total 0 to 99.9 LPM
Minute Volume (Spontaneous) 0 to 99.9 LPM
I:E Ratio 9.9:1 to 1:99
Peak Inspiratory Pressure -20 to 130 cm H2O
(On display)
Mean Airway Pressure -20 to 120 cm H20
(On display)
Plateau Pressure -20 to 130 cm H2O
(On display)
End Expiratory Pressure -20 to 99.9 cm H2O
(On display)
Delivered O2% 21 to 100%
Rapid, Shallow Breathing Index (f/Vt)0 to 500

Waveforms
Flow, Pressure and Volume Over Time

Loops
Single Flow, Pressure, Volume Waveform
Flow/Volume Loop and Pressure/Volume Loop
SaveableWaveform and Loops
Freeze Feature
Overlay Feature



Calculated Values (from maneuvers)

End Expiratory Pressure -20 to 120 cm H2O
(On display)
Expiratory Pause Pressure -20 to 120 cm H2O
(On display)
Auto PEEP -20 to 120 cm H2O


Alarms

Alarm Controls
Audible Alarm Volume Adjustable
Alarm Silence 120 sec.
Alarm Reset
High Inspiratory Pressure 10 to 105 cm H2O
Low Inspiratory Pressure 3 to 105 cm H2O
Low PEEP/CPAP Pressure 0 to 35 cm H2O
High Respiratory Rate 0 to 150 BPM
Low Exhaled Mandatory Tidal Volume 0 ml to 2500 ml
Low Exhaled Spontaneous Tidal Volume 0 ml to 2500 ml
Low Exhaled Minute Volume 0 to 60 LPM
High Exhaled Minute Volume 0 to 60 LPM
Apnea Interval 10 to 60 sec.
Alarm Status Indicators
  • Normal Indicator
  • Safety Valve Open
  • Ventilator Inoperative
  • Screen Locked
  • Alarm Silence
  • High Urgency Alarm
  • Medium/Low Urgency Alarm


Non-Adjustable Alarms
  • Circuit Occlusion
  • Low O2 Inlet Pressure
  • High Delivered O2%
  • Air Source Inoperative
  • Inspiratory Time Too Long
  • Low Delivered O2%


AC Power and Battery Indicators
  • Loss of Power (Audible)
  • Back-up Battery - Charging, In Use, Low
  • Loss of AC Power (Audible)
  • External Battery In Use
  • Mains


Self Test/Self Diagnostics
  • Short Self Test
  • Extended Self Test
  • Built In Support/Trouble Shooting for Service


Interface Ports
  • RS-232 Output and Nurse Call Output
  • Software Update/Upgrade
  • Via Serial Interface (Field updateable / upgradeable)


Dimensions
Height 16" (40.5 cm)
Width 14" (35.6 cm)
Depth w/ filter 24" (61.0 cm) 27.5" (69.9 cm)
Weight w/ back-up battery 66 lbs. (30 kg) 86 lbs. (39 kg)


Ventilator Mounting Options
  • Deluxe or Standard Cart
  • Wall Mount
  • Table Top
     
    Top
 
» Esprit Ventilator
» Home & Portable Ventilator
   - PLV® Continuum™
 
  Product Training:
  Invasive Ventilation  
 
 
Download Acrobat Reader
Click here to download Acrobat Reader
 
 
 

       
      Last modified: December 10, 2007
Copyright© 2003 - 2009 Mayo Healthcare Pty Ltd
Mayo Healthcare Pty Ltd & Mayo Hardware Pty Ltd are subsideries of Mayo Group Int. Pty Ltd

  Customer Service: 1300 360 226
Email: surgcserv@mayohealthcare.com.au