TCO2M® Transcutaneous CO2/O2 Monitor,
Model 860
|
|
|
| |
|
Technical
Specifications - TCO2M Monitor Model 860 |
 |
|
| Measurement
Range |
- Carbon Dioxide 0-255 mmHg (0-25.5 kPA). Oxygen 0-800 mmHg
(0-99.9 kPA)
|
| CO2 Metabolic Correction |
- Displayed pCO2 is compensated for metabolic
factors related to the temperature effect and CO2 production.
|
| Parameter
Alerts |
- CO2 and O2 limit alerts,
audible and visible. User selectable, continuously displayed,
retained
in memory.
|
| Alert
Silence |
- Two minute alert silence with LED indicator.
Audio Off with LED indicator (user can disable if needed).
|
| Temperature
Section |
- User selectable 37-45° C in 0.5° C increments.
- Can be continuously displayed.
- Audible & visible alarm
if actual/set temperatures differ by >0.2° C.
- Two
fully independent thermistors monitor and control temperature.
- Heater power shutdown with audible & visible alarm at
over-temp (45.5° C) or if thermistors differ by >0.5°.
|
| Site
Timer |
- Automatically tracks sensor site duration, provides audible
and visible alarms, automatic shutdown of heater power once
timer expires.
- User selectable durations in 0.5 hr increments
up to 5, 8, or 12 hour maximum, plus off.
|
| Local
Power |
- The power required to maintain sensor temperature
is measured and displayed.
- Range: 0-800 mW. User selectable
power reference point allows for quick power trend
analysis (LPR).
|
| Real
Time Trends |
- Continuously displayed user selectable/scalable real time
10, 20, 30 minute trends of CO2, O2 or
Local Power.
|
| Trend/Histogram
Memory |
- Battery backed 24 hour trends, 8 second resolution. Trend/Histogram
displays 30 min, 2, 4, 8, 12 hours.
|
| External
Devices |
- Standard RS232 serial outputs:
NovaCARD, computer, printer, others.
- Optional Analog Module, CO2,
O2 and LP.
|
| System |
- Power: 100-120/200-240
VAC, 50-60 Hz, 40 VA.
- Battery: Greater than 3 hour operation, 12 hour maximum recharge,
lead-acid gel-cell.
- Display Type: Cold Cathode Display (CCD), 5 × 1.5 inches,
W × H (12.7 × 3.8 cm). User adjustable contrast.
- Monitor Size: 3.3 × 9.0 × 8.0
inches, H × W × D
(8.4 × 22.8 × 20 cm).
- Weight: 8 pounds (3.6 kg).
- Temperature: Operating 10-40° C (50-104° F), Storage
-10-55° C (14-131° F).
- Relative humidity:
0-90%.
- Altitude/Pressure: Automatic barometric pressure
compensation.
- Range: 560-800 mmHg (71.9-106.6
kPa).
- Safety/Regulatory: Designed to comply with
EC and TUV (IEC 601-1), UL (2601), and
CSA (601-1-M90)
standards.
|
|
| |
|
Top |
| |
|
Technical
Specifications - Sensors |

|
|
| Parameter |
- Combination CO2/O2, CO2 only
or O2 only
|
| Principle
of Operation |
- Stow-Severinghaus pH/Clark-type polargraphic
|
| Response
Time (t90) |
|
| Stability |
|
| Diameter/Height |
|
|
| |
|
Top |
| |
|
Technical
Specifications - Gas Calibrator Model 868 |
|
|
| Calibration
System |
- Fully automatic using a true two-point sensor calibration
against known gases.
|
| Gas
Cylinders |
- Type: PD Contents: 16 liters at 16.5 bar, 21° C (0.6
cu. ft. at 240 psig, 70° F).
|
| Physical |
- Size: 3.3 × 9.0 × 16.5 in., H × W × D
(8.4 × 22.8 × 41.9 cm) includes cylinders.
- Weight: 3 lbs (1.4 kg)
|
| Power |
- All calibrator power provided by monitor.
|
| Diameter/Height |
|
|
| |
|
Top |
|
|
|