mayohealthcare logo  
 
Custom Search
       
 

Products

Ventilation and Respiratory Monitoring Group > Ventilation Monitoring Systems > Pulse Oximetry

Pulse Oximetry
OXYPLETH® Pulse Oximeter, Model 520A


    Technical Specifications

OXYPLETH® Pulse Oximeter, Model 520A
 
   Principle of Operation
  • Red/Infrared absorption.
    SpO2 (Oxygen Saturation)
  • Range: 0-100%
  • Accuracy: ±2% SpO2 (for 80-100% SpO2) (1 SD, or 68% of readings within claim) Unspecified for 0-79% SpO2
  • Resolution: 1%
  • Averaging: Menu selectable 2 or 8 seconds
   Pulse Rate
  • Range: 30-250 bpm
  • Accuracy: ±1% of full scale
  • Resolution: 1 bpm
  • Averaging: 8 seconds
   Sensors
  • Reusable Y-Sensor™ (can be sterilised and used with all patient populations) and reusable adult finger sensor.
   Plethysmogram
  • Pulsatile waveform with autogain, (autogain is menu selectable on/off).


    Top
    General 

OXYPLETH® Pulse Oximeter, Model 520A
 
   Alerts
  • Automatic and adjustable limits for SpO2 and Pulse Rate
  • Audio: Adjustable volume, 2 min. silence or OFF (LED indicators)
  • Visual: Flashing numerics upon violated limit(s) & red "Alert Bar". Indication of sensor disconnect, sensor off patient, low signal, insufficient light, light interference, pulse out of range, sensor faulty and monitor faulty conditions.
    Display
  • Type: Dot matrix, Cold Cathode Display (CCD)
  • Size: 5" W × 1.5" H (12.7 × 3.08 cm)
   Graphic Trend / Histogram
  • Memory: 24 hr., battery backed
  • Format: On-screen or printed, in 30 min., 2, 8, or 12 hr. segments
   Communication Output
  • Digital: RS232 (printer, NovaCARD, computer interface)
  • Analog: Optional module, 0-1VDC (numerics & waveforms)
   Physical
  • Size: 3.3" H × 9.0" W × 8.0" D (8.38 × 22.86 × 20.32 cm)
  • Weight: 7 lbs. 5 oz. (3.32 kg)
   Electrical
  • Power requirements: 100-120/200-240 VAC, 50-60 Hz, 30 VA
  • Battery: Sealed lead acid gel cell, 4 hr. life, 12 hr. recharge
   Environmental
  • Operating temperature: 50-104° F (10-40° C) Operating humidity: 0-90% relative (non-condensing)


    Top
 
» OXYPLETH®
» Models 512 &  513
» Model 515B &  Model 515C
 
NICO
Pulse Oximetry
Capnography
 
 
 
 
Download Acrobat Reader
Click here to download Acrobat Reader
 
Novametrix Logo
 
 

       
      Last modified: December 10, 2007
Copyright© 2003 - 2009 Mayo Healthcare Pty Ltd
Mayo Healthcare Pty Ltd & Mayo Hardware Pty Ltd are subsideries of Mayo Group Int. Pty Ltd

  Customer Service: 1300 360 226
Email: surgcserv@mayohealthcare.com.au