Model 515B & Model 515C Pulse Oximeter
|
|
|
| |
|
Technical
Specifications |

515B

515C
|
|
The Model 515B Pulse Oximeter has the same technical specifications
as the Model 515C excluding the Plethysmogram display.
| Principle
of Operation |
|
| SpO2 (Oxygen
Saturation) |
- Range: 0-100%
- Accuracy: ±2%
SpO2 (for 80-100% SpO2) (1 SD, or 68% of readings within claim)
Unspecified for 0-79%
SpO2
- Resolution: 1%
- Averaging: 8 seconds
|
| Pulse
Rate |
- Range: 30-250 bpm
- Accuracy: ±2
bpm
- Resolution: 1 bpm
- Averaging: 8 seconds
|
| Sensors |
- Reusable Y-Sensor™ (can be sterilized
and used with all patient populations) and reusable adult
finger sensor.
|
| Plethysmogram |
- Pulsatile waveform with autogain.
|
|
| |
|
Top |
| |
|
General |
| |
|
| Alerts |
- Automatic and adjustable limits for SpO2 and
Pulse Rate
- Audio: Adjustable volume, 2 min. silence or
OFF (LED indicators)
- Visual: Flashing numerics upon violated
limit(s) & red "Alert
Bar". Indication of sensor disconnect, sensor off
patient, low signal, insufficient light, light interference,
pulse out of range, sensor faulty and monitor faulty conditions.
|
| Display |
- Numerics: 7-Segment LED'sPulse Bar: LED
- Pulse Bar Graph proportional to signal strength
- Plethysmogram (Model 515C): Liquid Crystal Graphic module,
2.4" W × 1.3" H (6.0 x 3.3 cm)
|
| Physical |
- Size: 3.3" H × 9.0" W × 8.0" D
(8.38 × 22.86 × 20.32 cm)
- Weight: 6 lb (2.72
kg)
|
| Electrical |
- Power requirements: 100-120/200-240 VAC, 50-60 Hz,
30 VA
- Battery: Sealed lead acid gel
cell, 8 hr. life, 12 hr. recharge
Indicators for battery
operation, low battery, and
extremely low battery.
|
| Environmental |
- Operating temperature: 50-104° F (10-40° C)
- Operating
humidity: 0-90% relative (non-condensing)
|
|
| |
|
Top |
|
|
|