NICO (Non Invasive Cardiac Output)
NICO® Cardiopulmonary Management System


    Technical Specifications

NICO Cardiopulmonary Management System
 
Specifications for the NICO Cardiopulmonary Management System are listed for informational purposes only and are subject to change without notice.

   Cardiac Output
  • Measurement Frequency: Rebreathing cardiac output measurement made every three minutes, rebreathing period is 35 seconds.
  • Cardiac Output
    - Range: 0-19.9 liters/minute
    - Resolution: 0.1 liters/minute
  • Pulmonary Capillary Blood Flow (PCBF)
    - Range: 0-19.9 L/min
    - Resolution: 0.1 L/min
  • Cardiac Index
    - Range: 0-9.9 L/min/meter2
    - Resolution: 0.1 L/min/meter2
  • Stroke Volume
    - Range: 0-250 ml
    - Resolution: 1 ml
  • Rebreathing Valve/sensor:
    - Valve type: dual diaphragm, pneumatically controlled
    - Return spring: automatically returns valve to normal position
    - Resistance: 3cmH2O/L/min maximum
    - Rebreathed volume: normal position 35ml; rebreathing position 150-450ml
      (Standard size)
    - CO2/flow sensor: integrated into valve assembly
  • Parameter limits for NICO measurements:
    - VCO2: >20 ml/min
    - RR: >3, <60
    - Vt: >200
    - ETCO2: >15, <85 mmHg (<100 mmHg during rebreathing)
      > 2.0, <11.5 kPa or % (<13.5 kPa or % during rebreathing)
   CO2
  • Principle of Operation: Non-Dispersive Infrared (NDIR) absorption, dual wavelength ratiometric-single beam optics, mainstream sensor.
  • Response Time: Less than 60 ms
  • Gas composition effects: Operator selectable
  • CAPNOSTAT® CO2 Sensor:
    - Weight: Less than 18 g without cable
    - Sensor Size: 1.3 x 1.67 x .85 inches (3.3 x 4.2 x 2.2 cm), 8 foot cable (2.44 m)
    - Construction: Durable high performance plastic, ultra-flexible cable Shock   Resistant: Sensor will withstand a 6 foot drop to a tile floor
  • End Tidal CO2:
    - Range: 0-150 mmHg, 0-20 kPa or %
    - Accuracy: ± 2 mmHg for 0-40 mmHg, ± 5% of reading for 41-70 mmHg, ± 8% of   reading for 71-150 mmHg
  • Respiratory Rate:
    - Range: 1-150 breaths/min
    - Accuracy: ± 1 breath/min

   Flow Sensor
  • Flow Range (L/min) @ Pb 760 mmHg, room air, 35°C: 2 to 180
  • Flow Accuracy: Greater of ± 3% reading or .5 L/min
  • Minute Volume Range (L/min): 2 to 60
  • Tidal Volume Range (ml): 100 to 3000
  • Airway Pressure Range (cmH2O): -120 to 120

   SpO2
  • Oxygen Saturation
    - Range: 0-100%
    - Accuracy: ± 2% for 80-100% (for 1 standard deviation or approximately 68% of   readings), unspecified for 0-79%
    - Averaging Time: 2 seconds
  • Pulse Rate:
    - Range: 30-250 beats per minute
    - Accuracy: ± 1% of full scale (for 1 standard deviation or approximately 68% of   readings)
    - Averaging Time: 8 seconds

   Monitor
  • Classification (IEC601-1): Class I/internal power source, type BF, continuous operating mode
  • Operating Environment: 50-95° F (10-35° C), 0-90% relative humidity (non-condensing)
  • Size: Height 6.5 in., Width 10.75 in., Depth 9.5 in.
  • Weight: 10 pounds
  • Power: 100-240 Vac, 50-60 Hz, 70VA
  • Fuse Rating: 100-240 VAC, 0.5 A 250 V Slo-Blo (x2); 200-240 VAC, T 250 mA/250 V (x2)
  • Battery: Internal, Sealed lead-acid gel-cell, 45 minute life on full charge (on-screen life indicator), 12 hours recharge time.
  • Display: 4.625 x 3.5 inch EL, 320x240 pixels
  • Electromagnetic Emissions: Conforms to EMC Directive 89/336/EEC, CISPR Class A. Tested to EN55011 (1991) and CISPR11 (1990).
  • Electromagnetic Immunity: Conforms to EMC Directive 89/336/EEC, EN50082-1 (1992). Tested to IEC801-3 (1984) Radiated Immunity. Conforms to Medical Device Directive 93/42/EEC EN60601-1 (1992). Tested to IEC801-2 (1991) ESD, IEC801-4 (1988) EFT, and IEC1000-4-5 (1995) Surge Immunity.

   Communications
  • RS232 Communications Ports
  • Analog Input/Output Port (selectable, 0 to 1 volt range)

    Top
 
» NICO Cardiopulmonary
Management System
 
NICO
Pulse Oximetry
Capnography
 
 
 
 
Download Acrobat Reader
Click here to download Acrobat Reader
 
Novametrix Logo
 
 

       
      Last modified: December 10, 2007
Copyright© 2003 - 2009 Mayo Healthcare Pty Ltd
Mayo Healthcare Pty Ltd & Mayo Hardware Pty Ltd are subsideries of Mayo Group Int. Pty Ltd

  Customer Service: 1300 360 226
Email: surgcserv@mayohealthcare.com.au