TIDAL WAVE™ Hand-held Capnograph


    Technical Specifications

TIDAL WAVE™ Hand-held Capnograph

 
   Principle of Operation
  • Solid state mainstream sensor. Single beam, non-dispersive infrared (NDIR) absorption, ratiometric measurement.
    Initialisation Time
  • Capnogram within 5 seconds, full specifications within 60 seconds.
   Calibration
  • No calibration required.
   CAPNOSTAT® Mainstream CO2 Sensor
  • Size: 3.30 × 4.24 × 2.16 cm (1.30 × 1.67 × 0.85 in.)
  • Weight: 18 grams (0.64 oz.), cable excluded
  • Shock Resistance: Withstands repeated six foot drops to a hard surface.
   Airway Adaptors
  • Pediatric/Adult: Single patient use, added deadspace 5 cc.
  • Neonatal: Single patient use, added deadspace 0.5 cc
   CO2 (Carbon Dioxide)
  • Range: 0 - 100 mmHg
  • Accuracy:
    ±2 mmHg (for 0 - 40 mmHg),
    ±5% of reading (for 41 - 70 mmHg),
    ±8% of reading (for 71 - 100 mmHg)
  • Resolution: 1 mmHg
  • Response Time: 60 ms
  • Compensations: N2O & O2 (selectable), BP (automatic)
  • Graphics: Capnogram and 30 miute trend
  • Units: mmHg, kPa, or %
   Respiratory Rate
  • Accuracy: ±1 br/min
  • Range: 0-150 br/min
  • Resolution: 1 br/min

    Top
    Technical Specifications - General 

 
   No Respiration Alerts
  • Selectable: 20, 40, or 60 seconds
  • Audio: Two minute silence or audio off
  • Visual: On-screen timer and LED
    Display
  • Liquid Crystal Display
    - 28 × 64 pixels, 2.5 × 1.25 inches.
    - Adjustable contrast with backlight and plastic shield.
   Physical
  • TIDAL WAVE monitor
    - Size (maximum height, width, depth): 20 × 9.5 × 4.8 cm (8 × 3.75 × 1.875 in.)
    - Weight: 680 grams (24 oz.)
   Electrical
  • Battery
    - Rechargeable: Type DR30, Nickel-Metal-Hydride (NiMH): 4 hour life
    - Primary: High-Energy lithium cells, size AA (7 required): 5.5 hour life
  • External Supply (optional)
    - Input: 120 VAC 60 Hz, or 220 VAC 50 Hz
   Environmental
  • Operating temperature: 0 to +40 °C (32-104 °F)
  • Operating humidity: 0 to 90% RH non-condensing.


    Top
 
» TIDAL WAVE Sp
» TIDAL WAVE
» CAPNOGARD®
» CO2SMO®
» CO2SMO® Plus
 
NICO
Pulse Oximetry
Capnography
 
 
 
 