mayohealthcare logo  
 
Custom Search
       
 

Products

Ventilation and Respiratory Monitoring Group > Ventilation Monitoring Systems > Capnography

Capnography
TIDAL WAVE Sp™ Hand-held Capnograph/Pulse Oximeter


    Technical Specifications - Capnograph
TIDAL WAVE Sp™ Hand-held Capnograph/Pulse Oximeter   Model 715

TIDAL WAVE Sp™ Hand-held Capnograph/Pulse Oximeter  Options
 
   Principle of Operation
  • Solid state mainstream sensor. Single beam, non-dispersive infrared (NDIR) absorption, ratiometric measurement.
    Initialisation Time
  • Capnogram within 15 seconds, full specifications within 60 seconds.
   Calibration
  • No calibration required.
   CAPNOSTAT® Mainstream CO2 Sensor
  • Size: 3.30 × 4.24 × 2.16 cm (1.30 × 1.67 × 0.85 in.)
  • Weight: 18 grams (0.64 oz.), cable excluded
  • Shock Resistance: Withstands repeated six foot drops to a hard surface
   CO2 (Carbon Dioxide)
  • Range: 0 - 150 mmHg
  • Accuracy:
    ±2 mmHg (for 0 - 40 mmHg),
    ±5% of reading (for 41 - 70 mmHg),
    ±8% of reading (for 71 - 150 mmHg)
  • Resolution: 1 mmHg
  • Response Time: 60 ms
  • Units: mmHg, kPa, or %
  • Compensations: N2O & O2 (selectable), Pb (automatic)
  • Capnogram: Selectable sweep speed
   Respiratory Rate
  • Accuracy: ±1 br/min
  • Range: 0-150 br/min
  • Resolution: 1 br/min
   Sampling Flow Rate (Model 715 only)
  • 180 ml/min (for non-intubated patients)
  • Response time is less than 200 ms at 180 ml/min flow

    Top
    Technical Specifications - Pulse Oximeter








 
   Principle of Operation
  • Red/Infrared absorption
    SpO2 (Oxygen Saturation)
  • Range: 0-100%
  • Accuracy: ±2% SpO2 (for 80-100% SpO2)
    (1 SD, or 68% of readings within claim) Unspecified for 0-79% SpO2
  • Resolution: 1%
  • Averaging: Menu selectable 2 or 8 seconds
   Pulse Rate
  • Range: 30-250 bpm
  • Accuracy: ±1% of full scale
  • Resolution: 1 bpm
  • Averaging: 8 seconds
   Sensors
  • Reusable Y-Sensor (can be sterilized and used with all patient populations);
    reusable adult finger sensor

    Top
    Technical Specifications - General 

 
   Alerts
  • Limits:Adjustable limits for ETCO2, SpO2, Respiratory Rate and Pulse Rate. No Respiration Detected adjustable timeout.
  • Audio: Adjustable volume, two minute silence or off (LED indicators)
  • Visual: On-screen and LED
    Display
  • Type: Dot matrix, backlit (LCD)
  • Size: 2.2 × 1.1 inches
    Trending Capabilities
  • Memory: 24 hr., battery backed
  • Format: 30 minute on-screen or printed in trend or tabular mode.
    Download data to personal computer for use with NovaCARD software
    Communication Output
  • RS232 to Seiko DPU-414 printer, NovaCARD software or computer
   Physical
  • Size (height, width, depth): 20 × 9.5 × 4.8 cm (8 × 3.75 × 1.875 in.)
  • Weight: 793 grams (28 oz.).
   Electrical
  • External Supply (optional)
    - Input: 120 VAC 60 Hz, or 220 VAC 50 Hz
    - Battery: Nickel-Metal-Hydride (NiMH): 4.5 hour life
    - Battery: Standard AA lithium: 5.5 hour life
   Environmental
  • Operating temperature: 0 to +40 °C (32-104 °F)
  • Operating humidity: 0 to 90% RH non-condensing.


    Top
 
» TIDAL WAVE Sp
» TIDAL WAVE
» CAPNOGARD®
» CO2SMO®
» CO2SMO® Plus
 
NICO
Pulse Oximetry
Capnography
 
 
 
 
Download Acrobat Reader
Click here to download Acrobat Reader
 
Novametrix Logo
 
 

       
      Last modified: December 10, 2007
Copyright© 2003 - 2009 Mayo Healthcare Pty Ltd
Mayo Healthcare Pty Ltd & Mayo Hardware Pty Ltd are subsideries of Mayo Group Int. Pty Ltd

  Customer Service: 1300 360 226
Email: surgcserv@mayohealthcare.com.au