Products |
| Ventilation and Respiratory
Monitoring Group
> Ventilation Monitoring Systems > Capnography |
Capnography
CO2SMO® Mainstream
Capnograph & Pulse Oximeter
|
|
|
| |
|
Technical
Specifications - Capnograph |
 |
|
| Principle
of Operation |
- Solid state mainstream sensor. Single beam, non-dispersive
infrared (NDIR) absorption, ratiometric measurement.
|
| Initialisation
Time |
- Capnogram within 15 sec., full specifications within
60 sec.
|
| Calibration |
- Routine calibration is not required. 15
sec. adapter zero performed when changing to different
style of adapter.
|
| Accuracy
Verification |
- Accuracy verifier provided on sensor cable.
|
| CAPNOSTAT® Mainstream
CO2 Sensor |
- Size: 1.30 × 1.67 × 0.85 inches (3.30 × 4.24 × 2.16
cm)
- Weight: 0.64 oz. (18 grams) - cable excluded
- Shock Resistance: Withstands repeated 6 foot drops.
|
| Airway
Adapters |
- Adult/pediatric reusable or single patient
use (added dead space <5cc), neonatal reusable (added
dead space <.5cc), and sampling adapter for non-intubated
patients.
|
| Sampling
Flow Rate |
- 180 ml/min (for non-intubated patients).
|
| CO2 (Carbon Dioxide) |
- Compensations: N2O & O2 (selectable), BP (automatic)
- Capnogram: Selectable sweep speed (12.5 or 25 mm/sec)
- Range: 0 - 100 mmHg
- Accuracy:
±
2 mmHg (for 0 - 40 mmHg)
±
5% of reading (for 41 - 70 mmHg)
±
8% of reading (for 71 - 100 mmHg)
- Resolution: 1 mmHg
- Response Time: 60 ms
|
| Respiratory
Rate |
- Accuracy: ±1 br/min
- Range: 0-150 br/min
- Resolution: 1 br/min
|
|
| |
|
Top |
| |
|
Technical
Specifications - Pulse Oximeter |

|
|
| Principle
of Operation |
|
| SpO2 (Oxygen
Saturation) |
- Range: 0-100%
- Accuracy: ±2% SpO2 (for 80-100% SpO2) (1 SD, or
68% of readings within claim)Unspecified for 0-79% SpO2
- Resolution: 1%Averaging: Menu selectable, 2 or 8 sec.
- Audio: Pitch of pulse tone varies with SpO2 value.
|
| Pulse
Rate |
- Range: 30-250 bpm
- Accuracy: ±1% of full scale
- Resolution: 1 bpm
- Averaging: 8 sec.
|
| Sensors |
- Reusable Y-Sensor™ (can be sterilized
and used with all patient populations); reusable adult
finger sensor.
|
| Plethysmogram |
- Pulsatile waveform with autogain on/off selection.
|
|
| |
|
Top |
| |
|
Technical
Specifications - General |
 |
|
| Alerts |
- Limits: Adjustable limits for ETCO2, SpO2, Resp. and
Pulse Rates
- Audio: Adjustable volume, 2 min. silence or OFF (LED
indicators)
- Visual: On-screen & red "Alert Bar"
|
| Display |
- Type: Dot matrix, Cold Cathode Display (CCD)
- Size: 5" W × 1.5" H (12.7 × 3.08
cm)
|
| Graphic
Tremd / Histogram |
- Memory: 24 hr., battery backed
- Format: On-screen or printed, in 30 min., 2, 8, or
12 hr. segments
|
| Communication
Output |
- Digital: RS232 (printer, NovaCARD, computer interface)
- Analog: Optional module, 0-1VDC (numerics & waveforms)
|
| Physical |
- Size: 3.3" H × 9.0" W × 8.0" D
(8.38 × 22.86 × 20.32 cm)
- Weight: 8 lb (3.63 kg)
|
| Electrical |
- Power requirements: 100-120/200-240 VAC, 50-60 Hz,
30 VA
- Battery: Sealed lead acid gel cell, 2 hr. life, 12
hr. recharge
|
| Environmental |
- Operating temperature: 50-104° F (10-40° C)
- Operating humidity: 0-90% relative (non-condensing)
|
|
| |
|
Top |
|
|
|