mayohealthcare logo  
 
Custom Search
       
 

Products

Ventilation and Respiratory Monitoring Group > Ventilation Monitoring Systems > Capnography

Capnography
CO2SMO® Plus Respiratory Profile Monitor


    Technical Specifications - Capnograph

CO2SMO® Plus Respiratory Profile Monitor
 
   Principle of Operation
  • Solid state mainstream sensor. Single beam, non-dispersive infrared (NDIR) absorption, ratiometric measurement.
    Initialisation Time
  • Capnogram within 15 sec., full specifications within 60 sec.
   Calibration
  • Routine calibration is not required. 10 sec. adapter zero performed when changing to different style of adapter.
   Accuracy Verification
  • Accuracy verifier provided on sensor cable.
   CAPNOSTAT® Mainstream CO2 Sensor
  • Size: 1.30 × 1.67 × 0.85 inches (3.30 × 4.24 × 2.16 cm)
  • Weight: 0.64 oz. (18 grams) - cable excluded
  • Shock Resistance: Withstands repeated 6 foot drops.
   Airway Adapters
  • Adult/Pediatric reusable or single patient use CO2 adapters available separately or combined with flow sensor.
  • Reusable CO2 adapters also available.
   Sampling Flow Rate
  • Included for non-intubated patients (pump draws 180 ml/min).
   CO2 (Carbon Dioxide)
  • Compensations: Gas composition (selectable), Barometric pressure (automatic)
  • Capnogram: Selectable sweep speed
  • Range: 0 - 150 mmHg
  • Accuracy:
    ±2 mmHg (for 0 - 40 mmHg)
    ±5% of reading (for 41 - 70 mmHg)
    ±8% of reading (for 71 - 150 mmHg)
  • Resolution: 1 mmHg
  • Response Time: 60 ms
   Respiratory Rate
  • Accuracy: ±1 br/mi
  • Range: 0-150 br/min
  • Resolution: 1 br/min



    Top
    Technical Specifications - Flow Sensor

CO2SMO® Plus Respiratory Profile Monitor







 
   Principle of Operation
  • Fixed orifice, differential pressure (pediatric/adult and neonatal)
    Flow
  • Range: 0.25 - 25 L/min (neonatal), 2 - 180 L/min (pediatric/adult)
  • Accuracy:
    Greater of ±3% of reading or 0.125 L/min (neonatal), 0.5 L/min (pediatric/adult)
   Other (Range)
  • Minute Volume: 0.1 - 15 L (neonatal), 2 - 60 L (pediatric/adult)
  • Tidal Volume: 1 -500 mL (neonatal), 100 -3000 ml (pediatric/adult)
  • Airway Pressure: -120 to +120 cmH2O

    Top
    Technical Specifications - Pulse Oximeter

CO2SMO® Plus Respiratory Profile Monitor







 
   Principle of Operation
  • Red/Infrared absorption
    SpO2 (Oxygen Saturation)
  • Range: 0-100%
  • Accuracy:
    ±2% SpO2 (for 80-100% SpO2)
    Unspecified for 0-79% SpO2
    (1 SD, or 68% of readings within claim)
  • Resolution: 1%
  • Averaging: Menu selectable, 2 or 8 sec.
  • Audio: Pitch of pulse tone varies with SpO2 value
   Pulse Rate
  • Range: 30-250 bpm
  • Accuracy: ±1% of full scale
  • Resolution: 1 bpm
  • Averaging: 8 sec.
   Sensors
  • Reusable Y-Sensor™ (can be sterilized and used with all patient populations)
  • Reusable adult finger sensor
  • Single patient use sensors.

    Top
    Technical Specifications - General 

CO2SMO® Plus Respiratory Profile Monitor
 
   Parameters Measured
Parameters include ETCO2, SpO2, VCO2, mixed expired CO2, MV, VT, compliance, resistance, PIP, MAP, PEEP, (also identifies auto-PEEP), PIF, PEF, I:E, ti, te, rapid-shallow breathing index, and deadspace-related calculations: airway deadspace, alvolar minute ventilation, alveolar tidal volume, and with user entry of PaCO2: physiologic deadspace, alveolar deadspace, physiologic VD/VT.
   Trend Memory
All parameters trended for 24 hours, battery backed, available through serial output. VCO2, VT, and MV alv available on-screen.
    Display
  • Type: Dot matrix, Cold Cathode Display (CCD)
  • Size: 5" W × 1.5" H (12.7 × 3.08 cm)
   Alerts
  • Limits: Adjustable limits for ETCO2, SpO2, Respiratory Rate, Pulse Rate and No Respiration
  • Audio: Adjustable volume, 2 min. silence or OFF (LED indicators)
  • Visual: On-screen indicator and red "Alert Bar"
    RS232 Communication Output
  • Connects to a personal computer for printing and enhanced graphics and trending.
   Physical
  • Size: 3.3" H × 9.0" W × 8.0" D (8.38 × 22.86 × 20.32 cm)
  • Weight: 8 lb (3.63 kg)
   Electrical
  • Power requirements: 100-120/200-240 VAC, 50-60 Hz, 30 VA
  • Battery: Sealed lead acid gel cell, 2 hr. life, 12 hr. recharge
   Environmental
  • Operating temperature: 50-104° F (10-40° C)
  • Operating humidity: 0-90% relative (non-condensing)

    Top
 
» TIDAL WAVE Sp
» TIDAL WAVE
» CAPNOGARD®
» CO2SMO®
» CO2SMO® Plus
 
NICO
Pulse Oximetry
Capnography
 
 
 
 
Download Acrobat Reader
Click here to download Acrobat Reader
 
Novametrix Logo
 
 

       
      Last modified: December 10, 2007
Copyright© 2003 - 2009 Mayo Healthcare Pty Ltd
Mayo Healthcare Pty Ltd & Mayo Hardware Pty Ltd are subsideries of Mayo Group Int. Pty Ltd

  Customer Service: 1300 360 226
Email: surgcserv@mayohealthcare.com.au