mayohealthcare logo  
 
Custom Search
       
 

Products

Ventilation and Respiratory Monitoring Group > Ventilation Monitoring Systems > Capnography

Capnography
CAPNOGARD® Mainstream Capnograph


    Technical Specifications

CAPNOGARD® Mainstream Capnograph

 
   Principle of Operation
  • Solid state mainstream sensor. Single beam, non-dispersive infrared (NDIR) absorption, ratiometric measurement.
    Initialisation Time
  • Capnogram within 15 sec., full specifications within 60 sec.
   Calibration
  • Routine calibration is not required. 15 sec. adapter zero performed when changing to different style of adapter.
   Accuracy Verification
  • Accuracy verifier provided on sensor cable.
   CAPNOSTAT® Mainstream CO2 Sensor
  • Size: 1.30 × 1.67 × 0.85 inches (3.30 × 4.24 × 2.16 cm)
  • Weight: 0.64 oz. (18 grams) - cable excluded
  • Shock Resistance: Withstands repeated 6 foot drops
   Airway Adaptors
  • Adult/pediatric reusable or single patient use (added dead space <5cc)
  • Neonatal reusable (added dead space <.5cc)
  • Sampling adapter for non-intubated patients
   Sampling Flow Rate
  • 180 ml/min (for non-intubated patients).
   CO2 (Carbon Dioxide)
  • Compensations: N2O & O2 (selectable), BP (automatic)
  • Capnogram: Selectable sweep speed (12.5 or 25 mm/sec) Range: 0 - 100 mmHg
  • Accuracy:
    ± 2 mmHg (for 0 - 40 mmHg)
    ± 5% of reading (for 41 - 70 mmHg)
    ±8% of reading (for 71 - 100 mmHg)
  • Resolution: 1 mmHg Response Time: 60 ms
   Respiratory Rate
  • Accuracy: ±1 br/min
  • Range: 0-150 br/min
  • Resolution: 1 br/min


    Top
    Technical Specifications - General 

 
   Alerts
  • Limits: Adjustable limits for ETCO2 and Respiratory Rate
  • Audio: Adjustable volume, 2 min. silence or OFF (LED indicators)
  • Visual: On-screen & red "Alert Bar"
    Display
  • Type: Dot matrix, Cold Cathode Display (CCD)
  • Size: 5" W × 1.5" H (12.7 × 3.08 cm)
   Graphic Trend / Histogram
  • Memory: 24 hr., battery backed
  • Format: On-screen or printed, in 30 min., 2, 8, or 12 hr. segments
   Communication Output
  • Digital: RS232 (printer, NovaCARD, computer interface)
  • Analog: Optional module, 0-1VDC (numerics & waveforms)
   Physical
  • Size: 3.3" H × 9.0" W × 8.0" D (8.38 × 22.86 × 20.32 cm)
  • Weight: 8 lb (3.63 kg)
   Electrical
  • Power requirements: 100-120/200-240 VAC, 50-60 Hz, 30 VA
  • Battery: Sealed lead acid gel cell, 2 hr. life, 12 hr. recharge
   Environmental
  • Operating temperature: 50-104° F (10-40° C)
  • Operating humidity: 0-90% relative (non-condensing)
    Combination Capnograph /Pulse Oximeter Configuration
  • The CO2SMO® capnograph and pulse oximeter monitor has the same technical specifications as the CAPNOGARDand includes a pulse oximeter.


    Top
 
» TIDAL WAVE Sp
» TIDAL WAVE
» CAPNOGARD®
» CO2SMO®
» CO2SMO® Plus
 
NICO
Pulse Oximetry
Capnography
 
 
 
 
Download Acrobat Reader
Click here to download Acrobat Reader
 
Novametrix Logo
 
 

       
      Last modified: December 10, 2007
Copyright© 2003 - 2009 Mayo Healthcare Pty Ltd
Mayo Healthcare Pty Ltd & Mayo Hardware Pty Ltd are subsideries of Mayo Group Int. Pty Ltd

  Customer Service: 1300 360 226
Email: surgcserv@mayohealthcare.com.au