BiPAP® Synchrony™ Ventilatory Support System


    Specifications

Bipap Synchrony

Control Accuracy

Parameter Range Accuracy
IPAP 4 to 30 H2O ± 2 cm H2O
EPAP 4 to 30 H2O ± 2 cm H2O
CPAP 4 to 20 H2O ± 2 cm H2O
Breath Rate 0 to 30 BPM (PC and S/T) 4 to 30 BPM (T) Greater of 1 BPM or +/- 10% of setting when measured over a four minute period.
Timed Inspiration 0.5 to 3.0 sec ± (0.10% of setting + 0.1 sec)
Rise Time 100 to 600 msec ± 25%
Vent Ramp Disabled, 0.5, 1.0, 2.0, 3.0 cm H2O per breath (0.1+ 0.25 of setting) cm H2O/breath. Measured over the full duration of the programmed ramp.

  Physical
Dimensions 12" L x 7" W x 6" H
Weight 6 lbs

Electrical
AC Voltage 100 to 240 V, 50/60 Hz
DC Voltage 12 V
AC Current 1.25 A Maximum
DC Current 5.5 A Maximum

Trigger and Cycle
Spontaneous Trigger Shape Trigger
Volume 6 cc Above V Leak
Spontaneous Cycle Spontaneous Expiratory Threshold (SET)
Shape Cycle
IPAP Maximum of 3.0 sec
Flow Reversal

Mode of Operation
BiPAP Spontaneous, or Spontaneous/Timed, Timed, Pressure Control, and CPAP


    Top
 
» BiPAP Vision
» BiPAP FOCUS
» BiPAP Synchrony
» Cough Assist
» NIV Masks
 
  Product Training:
  Non-Invasive Ventilation  
 
 
Download Acrobat Reader
Click here to download Acrobat Reader
 
 
 

       
      Last modified: November 30, 2007
Copyright© 2003 - 2009 Mayo Healthcare Pty Ltd
Mayo Healthcare Pty Ltd & Mayo Hardware Pty Ltd are subsideries of Mayo Group Int. Pty Ltd

  Customer Service: 1300 360 226
Email: surgcserv@mayohealthcare.com.au