Specifications |
|
|
| Pressure: |
CPAP |
4 to 20 cm H2O (4 to 20 hPa) Resolution: 1 cm H2O (1 hPa) Dynamic Accuracy: ±5 cm H2O (±5 hPa) |
| |
EPAP |
4 to 20 cm H2O (4 to 30 hPa) Resolution: 1 cm H2O (1 hPa) Dynamic Accuracy: ±5 cm H2O (±5 hPa) |
| |
IPAP |
4 to 30 cm H2O (4 to 20 hPa) Resolution: 1 cm H2O (1 hPa) Dynamic Accuracy: ±5 cm H2O (±5 hPa) |
| Rise Time : |
Settings 1 to 6 (where 1 = 0.1 sec and 6 = 0.6 sec) Resolution: 1 Accuracy: ±10% of setting (at R=50, C=20) |
| Rate: |
1 to 30 BPM Resolution: 1 BPM Accuracy: ±1BPM or ±10 % of setting whichever is greater over a 4 minute period. I-Time and Rate settings cannot allow I-Time to exceed expiratory time. |
| I-Time: |
0.5 to 3 sec Resolution: 0.1 sec Accuracy: ±0.1+10% of setting sec I-Time and Rate settings cannot allow I-Time to exceed expiratory time. |
| Ramp Time: |
0 to 45 min Resolution: 5 min Accuracy: ±10% of setting |
| Ramp Start: |
4 cm H2O to EPAP (4 hPa to EPAP) Resolution: 1 cm H2O (1 hPa) |
Patient Parameters
| Patient Circuit Pressure: |
Range: 0 to 35 cm H2O (0 to 35 hPa) (continuous bar graph display) |
| Rate: |
Range: 0 to 99 BPM |
| Estimated Vt: |
Range: 0 to 4000 mL |
| Estimated MV: |
Range: 0 to 99 L/min |
| Patient Leak: |
Range: 0 to 200 L/min |
| Apnea Rate: |
Range: 0 to 99 per hour |
Alarms
| High System Pressure: |
5 cm H2O (5 hPa) over target IPAP pressure (Automatic) |
| Low System Pressure5: |
5 cm H2O (5 hPa) under target EPAP pressure (Automatic) |
| Apnea: |
Range: 20, 40, 60 sec or Off |
| Apnea Rate: |
Range: 5, 10, 20 apneas per hour or Off |
| Nurse Call: |
Supports normally open or normally closed systems |
Physical
| Dimensions: |
5.5" H x 11.4" W x 14" D (140 mm x 290 mm x 355 mm) |
| Weight: |
10 lb (4.5 kg) |
Power
| Input Rating: |
100VAC / 230VAC and 50/60 Hz |
| Battery Backup: |
3.8 amp-hour (NiMH) |
| Fuse: |
No replaceable fuse |
| Nurse Call: |
Rated current: 0.280 A Rated voltage: 250 V |
|
| |
|
|
| |
|