| Maximum Positive Pressure |
60 cm H2O (44 mm Hg) |
| Maximum Negative Pressure |
60 cm H2O (44 mm Hg) |
| Standards |
Conforms to UL STD 2601-1, certified to CAN/CSA - STD C22.2
No. 601.1-M90, CE 0413 |
| Maximum Inhalation Flow |
3.3 liters/second with inhale flow set to minimum; if set
to maximum inhalation, the flow is
the same as the exhalation flow |
| Maximum Exhalation Flow Capacity |
10 liters/second; actual flow depends on maximum pressure
and on airway resistance |
| Pressure Gauge |
-70 to 0 to +70 cm H2O; accuracy +/-2% of full scale |
| Mode of Operation |
Models CA-3000, CA-3200: Automatic and Manual Timing
Models CM-3000, CM-3200: Manual Timing |
| Inhalation, Exhalation, Pause Times |
CA-3000, CA-3200: Automatic mode, 0 to 5 seconds
CM-3000, CM-3200: User variable |
| “Off” Position |
Yes - connects to ambient |
| Blower Type |
Two-stage centrifugal blower with AC/DC universal motor |
| Input Voltage |
CA-3000, CM-3000: 100-120 VAC, 60Hz
CA-3200, CM-3200: 220-240 VAC, 50Hz |
| Input Power |
CA-3000, CM-3000: 300 VA
CA-3200, CM-3200: 600 VA |
| Dimensions |
292mm x 279mm x 419mm (HxWxD)
(11.5” x 11” x 16.5”) |
| Weight |
CA-3000, CA-3200: 11 kg (24 lbs)
CM-3000, CM-3200: 9.3kg (20.6 lbs) |
| Temperature Specifications |
Operating temperature range: +10 to +40°C (+50 to +104°F)
Storage temperature range: -20 to +50°C (-4 to +122°F) |
| Humidity Specifications |
Operating humidity range: 30% to 75%
Storage humidity range: 10% to 90% |